FREE: scheduling and accountability tools.COSTS EXTRA: test kits and laboratory or professional testing, bought separately.

FDA 510(k) record · Over the counter

Healstone Accurate Drug of Abuse Urine Test Cup

An FDA-cleared drug test cup. Clearance is verified; the FDA decision summary linked from the record contains no performance data.

Read this first

P in the Cup does not perform the testing or certify these products. The figures below come from one source document and describe one study design. They are not a general accuracy rating, and a home screen result is preliminary until a laboratory confirms it. See what accuracy figures mean.

Identity

Exactly what this record covers

Healstone Accurate Drug of Abuse Urine Test Cup: identity as stated in the source
FieldAs stated in the source
DeviceHealstone Accurate Drug of Abuse Urine Test Cup; Healstone Accurate Drug of Abuse Urine Test Cup Rx
ApplicantHealstone Biotech, Inc., Vancouver, Canada (correspondent: LSI International, Inc.)
510(k) numberK221871 (FDA database record)
DecisionSubstantially equivalent, decision date 30 August 2022 (received 28 June 2022); classification "Test, Amphetamine, Over The Counter" (product code NFT, 21 CFR 862.3100)
Intended useThe Accurate cup is for over-the-counter use; the Accurate cup Rx is for prescription use (applicant's 510(k) summary, page 3 of the summary / PDF page 9)
Sample typeHuman urine
Analytes and cutoffs (ng/mL)AMP 1,000 or 500; BUP 10; BAR 300; BZO 300; COC 300 or 150; EDDP 300; MET 1,000 or 500; MDMA 500; OPI/MOP 2,000 or 300; MTD 300; OXY 100; PCP 25; PPX 300; TCA 1,000; THC 50. "May contain any combination" (indications for use, PDF page 3)
PredicateHealstone Accurate Multi Panel Drug Urine Test Cup, K191841
Buy linkNo verified retailer link. No retailer listing has been matched to this exact device and SKU.
Last reviewed21 October 2026

Performance data

Published analytical data in linked review: not available

The FDA decision summary for K221871 (one page) says the 510(k) was reviewed under FDA's Triage-Quick Review Program and refers readers to the applicant's 510(k) summary. It contains no sensitivity, specificity, agreement or near-cutoff figures, so this record shows none.

The applicant's own 510(k) summary, linked from the same FDA record (42-page PDF), contains sections on precision, specificity, method comparison and a lay-user study (it describes 280 participants across two cup configurations, PDF page 38). We have not yet extracted or verified any figures from it, so none are published here. Until a reviewer completes that work, this record's performance tier is Unverified.

Limitations

  • Clearance verified, but the FDA decision summary is a one-page triage memo with no performance data; figures in the applicant's own 510(k) summary have not been extracted or verified by us.
  • Clearance applies to the device in this record. Retail or white-label cups need documentary proof that they are this exact device before this record applies to them.
  • Cutoffs vary by configuration (for example AMP 1,000 or 500 ng/mL); check the cutoff printed on the exact product.

Sources

General education only, not medical, legal or purchasing advice. Found an error? See the corrections log.

Page last updated