FREE: scheduling and accountability tools.COSTS EXTRA: test kits and laboratory or professional testing, bought separately.

Independent study · Professional point of care

Comparative analyses of three point-of-care urine drug test devices (J Appl Lab Med, 2022)

A peer-reviewed comparison of three professional point-of-care urine drug test devices against mass spectrometry. It is not a study of home kits.

Read this first

P in the Cup does not perform the testing or certify these products. The figures below come from one source document and describe one study design. They are not a general accuracy rating, and a home screen result is preliminary until a laboratory confirms it. See what accuracy figures mean.

Identity

Exactly what this record covers

Comparative analyses of three point-of-care urine drug test devices (J Appl Lab Med, 2022): identity as stated in the source
FieldAs stated in the source
StudyComparative Analyses of Three Point-of-Care Urine Drug Test Devices' Performance Characteristics for Use in Ambulatory Clinic Settings
AuthorsDerek C. Waggoner, Sarah R. Delaney, Nichole L. Korpi-Steiner
JournalThe Journal of Applied Laboratory Medicine, 2022, volume 7, issue 2, pages 495-502; DOI 10.1093/jalm/jfab095; PMID 34597363
Devices evaluatedProfile-V MEDTOX Scan drugs of abuse test; Quidel Triage TOX Drug Screen; Quidel Triage Rapid OXY-BUP-MDMA panel
SettingAmbulatory clinic (professional point-of-care) devices
Sample106 residual urine specimens; 84 of 106 (77%) positive for one or more drugs by quantitative mass spectrometry
Reference methodMass spectrometry, with manufacturer and reference-method cutoffs
AccessFull text for subscribers; we use only the publicly available abstract
Buy linkNo verified retailer link. No retailer listing has been matched to this exact device and SKU.
Last reviewed21 October 2026

Results stated in the abstract

Overall sensitivity and specificity, by device

These are overall figures, pooled across drug classes, using each device's manufacturer-stated cutoffs. The abstract says each device "had variable performance across each drug class"; the analyte-level results are in the subscriber-only full text and are not reproduced here.

Overall sensitivity and specificity versus mass spectrometry (abstract, PubMed 34597363)
DeviceMetricValueDenominatorReferenceTierSource
Profile-V MEDTOX Scan drugs of abuse testOverall sensitivity98.6%Not stated per device in the abstract (106 specimens in total; 84/106 positive for one or more drugs by mass spectrometry)Quantitative mass spectrometryAAbstract, Results (PubMed 34597363)
Profile-V MEDTOX Scan drugs of abuse testOverall specificity90.1%Not stated per device in the abstract (106 specimens in total; 84/106 positive for one or more drugs by mass spectrometry)Quantitative mass spectrometryAAbstract, Results (PubMed 34597363)
Quidel Triage TOX Drug ScreenOverall sensitivity97%Not stated per device in the abstract (106 specimens in total; 84/106 positive for one or more drugs by mass spectrometry)Quantitative mass spectrometryAAbstract, Results (PubMed 34597363)
Quidel Triage TOX Drug ScreenOverall specificity89%Not stated per device in the abstract (106 specimens in total; 84/106 positive for one or more drugs by mass spectrometry)Quantitative mass spectrometryAAbstract, Results (PubMed 34597363)
Quidel Triage Rapid OXY-BUP-MDMA panelOverall sensitivity100%Not stated per device in the abstract (106 specimens in total; 84/106 positive for one or more drugs by mass spectrometry)Quantitative mass spectrometryAAbstract, Results (PubMed 34597363)
Quidel Triage Rapid OXY-BUP-MDMA panelOverall specificity50%Not stated per device in the abstract (106 specimens in total; 84/106 positive for one or more drugs by mass spectrometry)Quantitative mass spectrometryAAbstract, Results (PubMed 34597363)

The authors conclude that the Profile-V MEDTOX Scan showed the best overall sensitivity and specificity of the three, that false positive and false negative results are possible, and that the best device may depend on the patient population and drugs of interest.

Limitations

  • Three professional point-of-care devices, 106 residual clinical specimens; we use only the abstract's pooled figures. Not evidence about any home kit.
  • Tier A applies only to these three devices, as tested in this study. It cannot be extended to other devices or to home kits.
  • Pooled figures hide drug-by-drug differences; per-device denominators are not given in the abstract.
  • Funding and conflicts of interest are not stated in the abstract and have not been checked in the full text.